B-565
WITHSTAND HIGH-VOLT HIGH-PRECISION INPUT
B-563SL-30
30CH SCREWLESS
INPUT TERMINAL
B-563
20CH SCREW
INPUT TERMINAL
B-563SL
20CH SCREWLESS
INPUT TERMINAL
Number of input channels 20ch 30ch 20ch 20ch
Input terminal type M3 screw-type Screwless M3 screw-type Screwless
Input method Photo MOS relay scanning system
All channels isolated, balanced input
Measurement ranges Voltage 20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50, 100 V, 1-5 V F.S.
Temperature Thermocouplesļ¼šK, J, E, T, R, S, B, N, C (W: WRe5-26)
RTD (only 3 wire type is supported) : Pt100, JPt100, Pt1000(IEC751) (* RTD cannot be connected.)
Temperature range: 100Ā°C, 500Ā°C, 2000Ā°C (In the case of Fahrenheit: 150Ā°F, 750Ā°F, 3000Ā°F)
Humidity 0 to 100% (Optional humidity sensor B-530 is required)
Maximum permissible input voltage Between +/ā€“ terminals 20mV to 2V range (60Vp-p)
5V to 100V range (110Vp-p)
Between input terminal/input terminal 600Vp-p 60Vp-p
Between input terminal/GND 300Vp-p 60Vp-p
Withstand voltage Between input terminal/input terminal 600Vp-p 350Vp-p怀1 minute
Between input terminal/GND 2300VACrms怀1 minute 350Vp-p怀1 minute
Voltage measurement accuracy Ā±(0.05%ofF.S.+10Ī¼V) Ā±0.1% of F.S.
Operating environment 0 to 45ā„ƒ
Dimensionsļ¼ˆ*including GL860ļ¼‰ 240Ɨ166Ɨ52.5mm (exclude protrusion) 240Ɨ158Ɨ52.5mm (exclude protrusion)
Weightļ¼ˆ*including GL860ļ¼‰ 1035g (exclude AC adapter & battery) 1010gļ¼ˆexclude AC adapter & battery)

B-565

< Between +/ā€“ terminals (A) >
ā€“ Maximum input voltage: 60Vp-p (Range of 20mV to 2V)
110Vp-p (Range of 5V to 100V)
< Between input terminal/input terminal (B) >
ā€“ Maximum input voltage: 600Vp-p
ā€“ Withstand voltage: 600Vp-p
< Between input terminal/GND (C) >
ā€“ Maximum input voltage: 300Vp-p
ā€“ Withstand voltage: 2300Vp-p at 1 minute

B-563SL-30 / B-563 / B-563SL

< Between +/ā€“ terminals (A) >
ā€“ Maximum input voltage: 60Vp-p (Range of 20mV to 2V)
怀怀怀怀怀怀怀怀 110Vp-p (Range of 5V to 100V)
< Between input terminal/input terminal (B) >
ā€“ Maximum input voltage: 60Vp-p
ā€“ Withstand voltage: 350Vp-p at 1 minute
< Between input terminal/GND (C) >
ā€“ Maximum input voltage: 60Vp-p
ā€“ Withstand voltage: 350Vp-p at 1 minute


GL860 TERMINALS

4 types of terminals including High Withstand Voltage B-565
are now available for the NEW GL860 Data Logger.

Select the terminal that best suits your measurement application.


B-565
Withstand High Voltage High Precision Input Terminal

B-563SL-30
30 Channel Screwless
Input Terminal

B-563
20 Channel Screw
Input Terminal

B-563SL
20 Channel Screwless
Input Terminal

Channel Expansion of Analog Signal Input Port

The GL860 is expandable up to 200 channels.

The input terminal and GL860 can be extended by using the connection cable for extension terminal (50cm / 2m)

*If the measurement is affected by noise, slow down the sampling interval.